Table 39. Physical Dimensions
Table 40. Power Requirements and Heat Output
Table 41. Temperature and Humidity Ranges
Height Width Depth Weight
50 cm (19.7-inch)
61 cm (23-inch), FPD
29 cm (11.4 inch), standard
34 cm (13.4 inch), cryo cooling
37 cm (14.6 inch), 6850 ALS
54 cm (21.3-inch)
64 cm (25.2-inch), auxiliary
EPC
< 23 kg (51 lb.)
Line voltage (±10%) Current Maximum power consumption Line frequency Heat output
Regular oven
100 V* (Japan) 15 A 1440 W 47.5 – 63 Hz
≤ 4800 BTU/hour
120 V* (USA) 15 A 1440 W 47.5 – 63 Hz
≤ 4800 BTU/hour
230 V* (Europe) 12 A maximum
8 A minimum
2000 W 47.5 – 63 Hz
≤ 4800 BTU/hour
Fast-heating oven
120 V* (USA) 20 A 2400 W 47.5 – 63 Hz
≤ 4800 BTU/hour
200 V* (Japan) 12 A 2400 W 47.5 – 63 Hz
≤ 4800 BTU/hour
230 V* (Europe) 11 A 2400 W 47.5 – 63 Hz
≤ 4800 BTU/hour
*Requires an isolated ground and dedicated outlet.
Temperature range Humidity range Altitude range
Operating 15–35°C 5–95%
Maximum 5–40°C 5–95%
≤ 2000 m