Verifying Performance
Software required
Released: March 2004 6850 Series II GC User Information page 54 of 256
Verifying Performance
This section contains Standard Operating Procedures along with typical
examples of test sample chromatograms.
Note that injection volumes listed do not necessarily indicate the absolute
volume injected. The volume reflects the graduation (plunger position) read
from a standard 10 µL syringe. For a heated inlet, the actual sample volume
injected may include an additional 0.4 to 0.7 µL, the volume of sample
volatilized from inside the syringe needle.
Note that the following procedures and results are intended to provide
evidence of a properly functioning inlet and/or detector system.
Software required
The procedures in this section require use of a ChemStation. They are written
for manual syringe injection.
If you are using Cerity Chemical, you can perform the chemical checkout for
each detector by loading the specified method and running the checkout
sample. Cerity Chemical cannot perform the noise test.
If a component is not yet supported on your data system version, you can enter
the appropriate chemical checkout settings using a control module.
TCD Checkout
Scope
Use this procedure to verify proper TCD operation with the split/splitless or
purged packed capillary column inlet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals