Automatic Liquid Sampler and Injector
Preparing sample vials
Released: March 2004 6850 Series II GC User Information page 218 of 256
3. Fill the solvent and waste bottles. See “Preparing the solvent and waste
bottles” on page 220.
4. Load the solvent bottles, waste bottles, and sample vials into the turret.
• 6850 Automatic Liquid Sampler—See “The sample turret” on page 229.
• 6850 Automatic Injector—See “Loading the turret” on page 236.
5. Load (or program) the GC sequence.
6. Run the GC sequence.
When the GC becomes ready, the sampler or injector begins the injections.
Preparing sample vials
Selecting sample vials
The sampler and injector turrets use clear or amber glass sample vials with
crimp-caps or screw-caps. Use amber glass vials for light-sensitive samples.
Refer to your Agilent catalog for consumables and supplies for acceptable vial
types. Incompatible sample vials cause turret errors.
Figure 36 shows the critical dimensions for sample vials and microvial inserts
to be used with the sampler and injector. These dimensions do not make up a
complete set of specifications.
Figure 36. Dimensions for sample vials and microvial inserts
Microvial insert
5 minimum
All dimensions in millimeters
Crimp cap sample vial
3.5
8.2 ± 0.2
34.5 mm
Maximum height of a vial with cap
11.7 ± 0.2
3.5
± 0.3
maximum
5 minimum
28.2 ± 0.3