Automatic Liquid Sampler and Injector
Syringes
Released: March 2004 6850 Series II GC User Information page 221 of 256
Empty and rinse each waste bottle after each multiple vial run. The syringe can
discard about 4 mL of waste into a waste bottle. See Figure 40.
Figure 40. Position of syringe tip when discarding waste
Syringes
Inspecting a syringe
Before installing a syringe:
Roll the syringe on the edge of a clean flat surface. If the tip of the needle
moves in a circle, straighten the shaft by bending it slightly near where it
connects to the syringe barrel and check it again. See Figure 41.
Figure 41. Syringe needle inspection
Check for a rough needle. The needle surface may have closely spaced
concentric ridges that act like a miniature file and abrade pieces of the septum
into the inlet or vial. The ridges are easy to see under 10X magnification.
If there are ridges, polish the needle by pulling it through a folded piece of fine
emery paper between your finger and thumb until the ridges are gone. Be
careful not to modify the tip of the syringe.
4 mL maximum waste
500 washes of a 10 µL syringe
1000 washes of a 5 µL syringe
200 washes of a 25 µL syringe
100 washes of a 50 µL syringe
50 washes of a 100 µL syringe
Capacity of one waste bottle
Bend here if necessary

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals