Site Prep Summary
Summary of GC benchtop and lab requirements
Released: March 2004 6850 Series II GC User Information page 255 of 256
Table 42. Gas Recommendations for Capillary Columns
Detector Carrier gas
Preferred makeup
gas
Second choice/
comments
Detector, anode purge, or
reference gas
Microcell elec-
tron capture
Hydrogen Argon/Methane Nitrogen Anode purge must be the same
as makeup
Helium Argon/Methane Nitrogen
Nitrogen Nitrogen Argon/Methane
Argon/
Methane
Argon/Methane Nitrogen
Flame ionization Hydrogen Nitrogen Helium Must use hydrogen and air as
detector gases
Flame photo-
metric
Hydrogen Nitrogen Hydrogen and air for detector
Helium Nitrogen
Nitrogen Nitrogen
Argon Nitrogen
Thermal
conductivity
Hydrogen Must be same as
carrier and
reference gas
Must be same as
carrier and
reference gas
Reference gas must be
same as carrier and
makeup
Helium
Nitrogen
Table 43. Gas Recommendations for Packed Columns
Detector Carrier gas Comments about carrier gas Detector, or reference gas
Microcell electron
capture
Nitrogen Maximum sensitivity Nitrogen
Argon/
Methane
Maximum dynamic range Argon/Methane
Flame ionization Nitrogen Maximum sensitivity Nitrogen
Helium maximum dynamic range
Flame photometric Hydrogen Hydrogen and air for detector
Helium
Nitrogen
Argon
Thermal conductivity Helium General use
Reference gas must be
same as carrier gas
Hydrogen
Maximum sensitivity
1
Nitrogen
Hydrogen detection
2
Argon
Maximum hydrogen sensitivity
2
1
Slightly greater sensitivity then helium. Incompatible with some compounds.
2
For analysis of hydrogen or helium. Greatly reduces sensitivity for other compounds.