Messages
Warning messages
Released: March 2004 6850 Series II GC User Information page 85 of 256
WARNING 107 Det config changed
WARNING 109 Inlet config changed
WARNING 111 Column config changed
WARNING 113 Aux 3 config changed
WARNING 114 Aux 4 config changed
WARNING 115 Aux 5 config changed
Possible causes: There is a new detector/inlet/column/auxiliary EPC module in the GC
The detector/inlet/column/auxiliary EPC module in the GC does not match the method
The method was changed to default detector/inlet/column/auxiliary EPC
module values because either a different one was installed, or a newly loaded
method uses a different detector/inlet/column/auxiliary EPC module than is
available.
If a new detector/inlet/column/auxiliary EPC module, the GC will enter default
setpoints for the new type.
If the method uses a different detector/inlet/column/auxiliary EPC module, the
method is unchanged and the warning message flashes.
WARNING 116 Run log full
The Run log can hold 50 entries of information. When it reaches the maximum,
this warning will appear.
WARNING 117 Inlet calib deleted
WARNING 119 Det calib deleted
WARNING 121 Aux calib deleted
Possible cause: Power failure during recalibration
While entering a new flow module calibration, the GC crashed or encountered
a power failure before the recalibration was complete. The warning appears
when the instrument is turned on again.
The calibration for the module returned to the default. Recalibrate.
WARNING 128 Inlet flow cal failed
Possible cause: Defective or leaking flow module
An attempted auto calibration of the flow sensor offset for the inlet exceeded
the allowable calibration range. The previous calibration setpoints remain in
memory.
WARNING 130 Aux cryo disabled
Possible cause: Incorrectly installed inlet cryo valve or valve box
The inlet cryo valve is also being used by aux thermal zone 1. The valve is
disabled.