Automatic Liquid Sampler and Injector
Faults
Released: March 2004 6850 Series II GC User Information page 241 of 256
Align Mode light is on, all others are flashing
(6850 Automatic Liquid Sampler only)
Probable cause
Turret alignment is in progress.
Suggested action
Install the calibration probe and press the Align button to complete the
alignment.
All lights are on
Probable cause
The cables connections are loose
There is a board failure
There is a firmware revision conflict
Suggested action
1. Check all cable connections.
2. Turn the instrument off, then on again.
3. If the lights remain on, obtain Agilent service.