EasyManuals Logo

Agilent Technologies 6850 Series II User Manual

Agilent Technologies 6850 Series II
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Verifying Performance
FPD Checkout
Released: March 2004 6850 Series II GC User Information page 77 of 256
Background offset checks
1. If lit, turn the FPD flame off.
2. Display the signal output to determine flame-off background offset. A
stable value of < 40 display units is acceptable.
3. Ignite the flame. You should see an increase in detector output to a stable
value.
If not, confirm ignition by holding a mirror or shiny surface near the
aluminum exhaust tube, with the rubber drip tube removed, and observing
condensation if the flame is lit.
If the flame has not ignited, see “Flame ignition problems” on page 208.
Final time 2 0 min
Purge time 0.75 min
Purge flow 60 mL/min
Detector
Temperature 200°C
H
2
flow 75 ±2 mL/min
Air flow 100 ±2 mL/min
Makeup flow (N
2
) 62mL/min
Offset, flow off (
0
-fa) Should be < 40 display units
Offset, flame on (
0
+fb) < [(
0
-fa) + 85 display units]
Oven
Initial temp 60°C
Initial time 0 min
Rate 1 25°C/min
Final temp 110°C
Final time 0 min
Rate 2 10°C/min
Final temp 170°C
Final time 3 min
Table 20. FPD Checkout Conditions (Method FPDCKO) (Continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals