114 Operation Manual
6 General Maintenance
To Disassemble the EI Ion Source
To Disassemble the EI Ion Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
1 Remove the ion source. See “To Change the Ion Source” on
page 94.
2 Remove the filaments. See “To Remove a Filament” on
page 126.
3 Separate the ion focus and insulator assembly from the
source body. (See Figure 29.)
4 Remove the two screws from the ion focus insulator and
separate it from the ion focus lens.
5 Remove the extractor lens and its insulator from the source
body.
6 Remove the two screws from the RIS bayonet mount on the
bottom of the source body, and remove the RIS bayonet
mount.
7 Remove the repeller assembly from the source body.
8 Disassemble the repeller assembly by separating the repeller,
the repeller cap, ring heater and sensor assembly. (See
Figure 29.)
9 Remove the clocking button from the repeller cap by
removing the screw that holds it in place. (See Figure 29.)

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals