104 Operation Manual
5 Using the Removable Ion Source
To Remove the RIS Probe Extraction Tool from the Instrument
To Remove the RIS Probe Extraction Tool from the Instrument
This procedure is part of an automated series of steps
controlled by MassHunter. Do not attempt this procedure unless
prompted by MassHunter. The RIS Status area on the GC/Q-TOF
Tune tab must show the Gate Valve is Open and the RIS vacuum as
0.000 Torr.
1
Undo the latch on the right side of the RIS probe extraction
tool and open the door.
2 Lift the RIS probe extraction tool up to release it from its
hinges.
3 Replace the RIS probe extraction tool with the RIS chamber
door.
4 Line up the hinge pins on the RIS chamber door with the
slots on the instrument hinge, and lower the pins into the
hinge.
5 Swing the RIS chamber door closed and secure the door with
the latch on the right side.
6 Return to step 16 in the “To Change the Ion Source” on
page 94.
WARNING
The 7200 Q-TOF GC/MS operates at high temperatures. Make sure
the RIS cooling chamber is at a safe temperature before attempting
to remove it from the instrument.