Periodic maintenance
As listed in Table 9, some maintenance tasks for the vacuum
system must be performed periodically. These include:
• Checking the foreline pump fluid color and level (every week)
• Checking the calibration vial(s) (every 6 months)
• Replacing the foreline pump oil and oil mist filter (every 6
months or as needed)
Failure to perform these tasks as scheduled can result in
decreased instrument performance. It can also result in damage
to your instrument.
Other procedures
Tasks such as replacing an ion vacuum gauge should be
performed only when needed by a certified Agilent service
personnel. See the Agilent 7200 Accurate-Mass Q-TOF GC/MS
Troubleshooting and Maintenance Manual and see the online
help in the MassHunter WorkStation software for symptoms
that indicate this type of maintenance is required.
More information is available
If you need more information about the locations or functions of
vacuum system components, see the Agilent 7200
Accurate-Mass Q-TOF GC/MS Troubleshooting and Maintenance
Manual.
Most of the procedures in this chapter are illustrated with video
clips on the Agilent GC/MS Hardware User Information &
Instrument Utilities and 7200 Q-TOF GC/MS User Information
disks.