52 Operation Manual
3 Operating in Electron Impact (EI) Mode
To Set Monitors for Temperature and Vacuum Status
To Set Monitors for Temperature and Vacuum Status
A monitor displays the current value of a single instrument
parameter. They can be added to the standard instrument
control window. Monitors can be set to change color if the
actual parameter varies beyond a user-determined limit from its
setpoint.
Procedure
1 In the Instrument Control panel of the MassHunter Workstation
Acquisition software, click on the Method icon.
2 Select Method > Edit Monitors to display the Select Monitors
dialog box. See Figure 11.