Operating in Electron Impact (EI) Mode 3
To Calibrate the Column
Operation Manual 59
parameters listed (temperature, inlet and outlet pressures,
and gas type) are those used in the method to determine the
holdup time. Change any parameters that are different than
those used in your method.
11 When the new column length appears, click OK to save the
changes.
12 Click OK on the Calibrate Columns screen to save the
calibration.
Figure 15 Calculate Column Length dialog box