86 Operation Manual
4 Operating in Chemical Ionization (CI) Mode 
To Operate the Reagent Gas Flow Control Module 
2 Use the parameters in the CI Reagent Gas Control area to 
control reagent gas flow.
CI Reagent - Selects the reagent gas from the dropdown menu.
CI Gas Flow - Enter percent of maximum volumetric flow for 
the selected reagent gas. The actual % transmitted by the flow 
controller is shown next to this entry. 20% is a good value for 
PCI and 40% is a good value for NCI.
Set button - Opens the selected reagent gas supply valve and 
controls the reagent gas flow to the entered setpoint.
Off button - Turns off the reagent gas flow.
Purge button - Opens the selected reagent gas valve for 6 
minutes to clear the system of unwanted compounds.
Pumpout button - Closes the reagent gas valve for 4 minutes 
and evacuates the system of reagent gases. At the end of the 
pumpout time the selected reagent gas valve opens.