86 Operation Manual
4 Operating in Chemical Ionization (CI) Mode
To Operate the Reagent Gas Flow Control Module
2 Use the parameters in the CI Reagent Gas Control area to
control reagent gas flow.
CI Reagent - Selects the reagent gas from the dropdown menu.
CI Gas Flow - Enter percent of maximum volumetric flow for
the selected reagent gas. The actual % transmitted by the flow
controller is shown next to this entry. 20% is a good value for
PCI and 40% is a good value for NCI.
Set button - Opens the selected reagent gas supply valve and
controls the reagent gas flow to the entered setpoint.
Off button - Turns off the reagent gas flow.
Purge button - Opens the selected reagent gas valve for 6
minutes to clear the system of unwanted compounds.
Pumpout button - Closes the reagent gas valve for 4 minutes
and evacuates the system of reagent gases. At the end of the
pumpout time the selected reagent gas valve opens.

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals