54 Agilent 7890B Installation
1 Installing the GC
line thermal zone will be labeled. If configured, skip to
step 2. If unconfigured, continue with the steps below.
b Press [Options], select Keyboard & Display and press [Enter].
c Scroll down to Hard Configuration Lock and press [Off/No].
d Press [Config][Aux Temp #]. Scroll to the aux thermal zone
for the MS transfer line and press [Enter].
e With the cursor on the Install Heater line, press [Enter]. A
caution appears. Press [Clear].
f Press [Mode/Type], scroll to the correct transfer line type
and press [Enter]. A caution appears. Press [Clear].
g When prompted to reboot, perform a reboot from the GC
keyboard. Press [Options], scroll to Communications, press
[Enter], scroll to Reboot GC? and press [On/Yes] twice.
2 Configure the MS on the GC.
a Press [Config], scroll to Mass Selective Detector, and press
[Mode/Type].
b On Unconfigured:, press [Mode/Type], then press [Enter].
c When prompted, perform a reboot from the GC keyboard.
3 Select the MS type. (Note that you can only enable direct
communication when launching MassHunter if an
instrument that supports enhanced communication is
selected here.)
a Press [MS/Aux Det].
b Scroll to MS type and press [Mode/Type].
c Select the MS model from the list, then press [Enter].
4 Set MS hardware options for Parts Finder. If your installed
source or pump is not listed, leave that item unconfigured.
This data helps Parts Finder present parts data specific to
your system, so a blank field means only that Parts Finder
will show all available choices.
a Press [MS/Aux Det].
b Scroll to Source, then press [Mode/Type].
c Select the source type, then press [Enter].
d Scroll to the other entries, and use the [Mode/Type] key to
select the appropriate options. Available entries include:
Second source, HV Pump, and Rough pump. For MS
instruments with only one source, set Second source to NOT
PRESENT.
e Scroll to Serial # and input the MS serial number. Press
[Enter].

Table of Contents

Other manuals for Agilent Technologies 7890B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890B and is the answer not in the manual?

Agilent Technologies 7890B Specifications

General IconGeneral
BrandAgilent Technologies
Model7890B
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals