26 Agilent 8453 Dissolution Testing System Installing and Operating Manual
2Setting Up
Setting Up Your Sampling System
Dissolution Bath
A dissolution probe kit and a packet of 40 filter tips are part of the valve and
multicell transport sampling systems.
A dissolution probe kit consists of eight fittings with a probe (25 cm, 1/8-inch
od) and union attached.
1 Cut the probes to a suitable length for your dissolution bath, if necessary.
2 Attach a filter to each probe.
3 Insert the probes into your dissolution vessels.
Attach the necessary fitting and tubing to the union, see “Connecting Multicell
System Tubing” on page 31, and “Connecting Valve System Tubing” on
page 49.
Figure 1 Probe for Dissolution Bath
Union
Fitting
Probe
Filter
NOTE
If you are using other filters with smaller pore size than 45 µm they may reduce the flow.
Some filters, if not washed before use, release monomers that pollute the flow cell
windows and or cause air bubbles to stick in the flow cells.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals