66 Agilent 8453 Dissolution Testing System Installing and Operating Manual
2Setting Up
Installing Your Multibath System
3 Connect the flow cell to the pump, see Figure 6 on page 32. Connect the
0.5 m tubing with the short cell fitting to the outlet of the flow cell, and the
fitting to a union. Connect a conical adapter to the union. Insert the conical
adapter into the pump tubing on the inlet side of the pump.
4 Connect the valve unit to the flow cell. Connect the 1.4 m or 2.2 m tubing
with the long cell fitting to the inlet (marked with
) of the flow cell.
Connect the valve fitting to the valve outlet.
5 Connect pump to waste. Insert the conical adapter into the pump tubing on
the outlet side of the pump and put the transparent fitting into a waste
beaker.
6 Repeat step 1 through step 5 for the other baths.
Replacing Tubing
General instructions for attaching new fittings to tubing are provided in a
replacement PTFE tubing kit (part number 5041-2191). This kit contains 10 m
of PTFE tubing.
NOTE
To minimize carry over, use the 1.4 m tubing.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals