Agilent 8453 Dissolution Testing System Installing and Operating Manual 27
Setting Up 2
Installing a Multicell-transport-based Sampling System
Installing a Multicell-transport-based Sampling System
The multicell-transport-based sampling system samples from one bath online.
The system uses a multicell transport with eight flow cells—one cell for the
blank, up to six cells for the individual vessels and one cell for a control—and a
multichannel pump to sample all vessels simultaneously. The dissolution
medium is recycled so that there is no change due to the sampling system in
medium volume during the test.
To install the multicell system you will need the following parts:
• 8-Position multicell transport
• Peristaltic pump 8VS
• Multicell tubing kit
• Flow cells, 1 mm path length (8)
• Peristaltic pump–remote cable
• Dissolution probe kit
Figure 2 Multicell-transport-based Sampling System (G1127A)
Peristaltic pump
remote cable
Spectrophotometer
Multicell transport
Flow cell
Tubin g
Vessel 1
Bath
Peristaltic
pump 8VS
Control vessel
Blank vessel