EasyManuals Logo

Agilent Technologies 8453 User Manual, Installing And Operating Manual

Agilent Technologies 8453
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
28 Agilent 8453 Dissolution Testing System Installing and Operating Manual
2Setting Up
Installing a Multicell-transport-based Sampling System
Installing Your Eight-position Multicell Transport
To install your eight-position multicell transport in the Agilent 8453
spectrophotometer, refer to your Multicell Transport for Agilent 8453
UV-visible Spectrophotometer Operator’s Manual.
Installing Your Peristaltic Pump 8VS
Each peristaltic pump 8VS is identified by a serial number on a label at the
rear. Write down the serial number of your pump here:
_____________________________________________________________________
Installing the pump comprises three stages:
“Stage 1: Installing the Cables” on page 28,
“Stage 2: Installing the Pump Tubing” on page 29, and
“Stage 3: Installing the Tubing Cassette” on page 30.
Stage 1: Installing the Cables
The voltage selector and cable connectors are at the rear panel.
1 Set the correct voltage option (115 V or 230 V) for operation from your line
power supply.
2 Check the correct fuse is installed.
3 Push connector of power cord into line-power socket.
NOTE
Position 1 is at the left side of the multicell transport.
Position 8 is at the right side of the multicell transport.
WARNING
Make sure line-power cord is disconnected before changing line-voltage setting.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals