70 Agilent 8453 Dissolution Testing System Installing and Operating Manual
3 Using the Peristaltic Pumps 1VS and 8VS
Installing Your Pump
Installing Your Pump
The peristaltic pumps 1VS and 8VS are for one and eight pumping channels.
The pumps are designed for liquid samples. Due to the peristaltic pump
principle only the tubing is in contact with the liquid.
The pumps can be operated manually and remotely via a digital interface. If
used for automated sampling in the dissolution testing application, only the
remote operation is applied.
Table 1 on page 15 shows the Physical Specifications Peristaltic Pumps 1VS
and 8VS and Table 2 on page 16 shows the Performance
Specifications Peristaltic Pumps 1VS and 8VS.
The pump operates on 100–120 V AC or 220–240 V AC and delivers flow up to
14 ml/min and 2 bar from each channel. The pump has a pressure-wheel
assembly for one or eight tubing-cassettes. The pump can be remotely
controlled to start and stop, and change direction, or it can be operated
manually from the front panel.
Viewing the Pump
The pump has three functional parts:
controls,
cable connectors, and
cassettes.
Controls
The front panel controls are a line power switch, a three-position control
switch, a flow control switch, and a safety symbol. See Figure 28.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals