EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>8453

Agilent Technologies 8453 User Manual, Installing And Operating Manual

Agilent Technologies 8453
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
90 Agilent 8453 Dissolution Testing System Installing and Operating Manual
4 Using the Valve Unit and Valve-pump Controller
Troubleshooting Your Valve System
Contamination Between Ports
This may be caused by too low flow rate or flow restrictions, that is, the
previous sample has not been flushed entirely from the tubing or flow cell, and
by a defective valve.
To check for Flow Restrictions—Too Low Flow:
1 Check for flow restrictions, for example blocked filter tips, or for bends in
the tubing which may restrict flow.
2 Set the flow control to a higher rate or longer pumping time if necessary.
3 Check for dirty tubing. If the tubing is dirty, see “Cleaning Tubing and
Valve on page 91.
To check for a Defective Valve:
This check uses pure solvent in the odd-numbered channels (1, 3, 5, and 7) and
a sample solution of about 1 AU in the even-numbered channels (2, 4, 6, and
8). You measure the absorbance in the odd-numbered channels where the pure
solvent should give no difference in absorbance compared to the solvent
reference, if there is no cross-contamination between two adjacent valve ports.
You then repeat the check using sample solution in the odd-numbered
channels and pure solvent in the even-numbered channels.
1 Measure a blank on air, that is no flow cell in the sample area, for an
instrument zero.
2 Measure the absorbance of solvent directly in your flow cell, that is without
passing through the valve. Use this measurement as a solvent reference
during the next steps.
3 Prepare a sample solution, that consists of an absorbing substance
dissolved in the same solvent to give an absorbance of about 1.0 AU.
4 Place the tubing for valve positions 1, 3, 5, and 7 in the solvent you
previously used for the reference.
5 Place the tubing for valve positions 2, 4, 6, and 8 in the sample solution.
6 Pump the solvent and sample solution into the flow cell from each valve
position and measure the absorbance of channels 1, 3, 5, and 7 (solvent
reference). Make sure the solvent and sample solution reach the flow cell
for your selected pumping time. The pumping time depends on the length of
tubing.
7 Compare the results with the solvent reference. An absorbance of not more
than 0.005 AU for the solvent indicates less than 0.5% cross contamination.

Table of Contents

Other manuals for Agilent Technologies 8453

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8453 and is the answer not in the manual?

Agilent Technologies 8453 Specifications

General IconGeneral
BrandAgilent Technologies
Model8453
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals