EasyManua.ls Logo

Agilent Technologies 8648A - Page 302

Agilent Technologies 8648A
481 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-46
Adjustments
Agilent Service Support Software
1. Select the Device Type of the test equipment to be removed (15).
2. Select the serial number of the test equipment to be removed from the Equipment field
(17)
3. Click Remove (18).
4. Click Close (19).
Editing Test Equipment
Editing test equipment parameters is accomplished using the Test Equipment dialog box.
1. Select the Device Type of the test equipment to be edited (15).
2. Select the model of the test equipment to be edited from the Models field (16).
3. Select the serial number of the test equipment to be edited (17).
4. Click Edit (20). Change the field parameters as necessary.
5. Click OK (12).
6. Click Close (19).
Adding Device Drivers
CAUTION The following three procedures: Adding/Removing Device (Test Equipment)
Drivers, Adding/Removing Test Drivers, and Adding/Removing Datapacks
are included for informational purposes. These procedures should only be

Table of Contents

Other manuals for Agilent Technologies 8648A

Related product manuals