Service Guide 10-7
8719ET/20ET/22ET Service Key Menus and Error Messages
8719ES/20ES/22ES Service Menus - Internal Diagnostics
RF Network Analyzers
selects either normal (NORM) or Special (SPCL) (tighter)
limits for the Operator’s Check. The SPCL limits are
useful for a guard band.
(POWLLIST) accesses the following Edit List menu to allow
modification of the external power loss data table.
accesses the power loss/sensor lists menu:
• selects the A or B power sensor calibration factor
list for use in power meter calibration measurements.
• (CALFSENA) accesses the Edit List menu
to allow modification of the calibration data table for power sensor A.
• (CALFSENB) accesses the Edit List menu
to allow modification of the calibration data table for power sensor B.
• (POWLLIST) accesses the Edit List menu to allow
modification of the external power loss data table that corrects
coupled-arm power loss when a directional coupler samples the RF
output.
generates printed graphs of verification results when
activated during a system verification.
Edit List Menu
To access this menu, press
and then press one of the following: or
or .
selects a segment (frequency point) to be edited, deleted
from, or added to the current data table. Works with the
entry controls.
(SEDI[D]) allows modification of frequency, cal factor and loss values
previously entered in the current data table.
(SDEL) deletes frequency, cal factor and loss values previously
entered in the current data table.
(SADD) adds new frequency, cal factor and loss values to the
current data table up to a maximum of 12 segments
(frequency points, PTS).
(CLEL) deletes the entire current data table (or list) when is
pressed. Press to avoid deletion.
(EDITDONE) returns to the previous menu.