EasyManua.ls Logo

Agilent Technologies 8719ET - Adjusting the Low Band Power; Adjusting the High Band Power (8722 ET;ES)

Agilent Technologies 8719ET
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Guide 3-21
8719ET/20ET/22ET Adjustments and Correction Constants
8719ES/20ES/22ES Output Power Adjustments
RF Network Analyzers
3. Press .
4. To switch the power DAC on, press the following:
5. Adjust the mid band power offset resistor (A59 R71) for a power meter reading that is
equal to the value (±0.1 dB) listed under “Power A” in Table 3-3.
6. Press .
7. Adjust the mid band power slope resistor (A59 R88) for a power meter reading that is
equal to the value (±0.1 dB) listed under “Power A” in Table 3-3.
8. The adjustments in step 5 and step 7 interact with each other, so repeat both
adjustments until both are in specification.
Adjusting the Low Band Power
1. Press .
2. Adjust the low band power offset resistor (A59 R70) for a power meter reading that is
equal to the value (±0.1 dB), from the “Power A” column in Table 3-3.
3. Press .
4. Adjust the low band power slope resistor (A59 R86) for a power meter reading that is
equal to the value (±0.1 dB), from the “Power A” column in Table 3-3.
Adjusting the High Band Power (8722ET/ES)
1. Press .
2. Adjust the high band power offset resistor (A59 R80) for a power meter reading that is
equal to the value (±0.1 dB), from the “Power A” column in Table 3-3.
3. Press .
4. Adjust the high band power slope resistor (A59 R87) for a power meter reading that is
equal to the value (±0.1 dB), from the “Power A” column in Table 3-3.
Sweep Setup
CW FREQ
2.56
G/n
System
SERVICE MENU
SERVICE MODES
SRC ADJUST MENU
POWER DAC[ON]
1000 x1
Sweep Setup
CW FREQ
20 G/n
50 M/µ
2.3 G/n
20.1 G/n
40 G/n

Table of Contents

Other manuals for Agilent Technologies 8719ET

Related product manuals