14-28 Service Guide
Assembly Replacement and Post-Repair Procedures 8719ET/20ET/22ET
Removing the A3 Disk Drive Assembly 8719ES/20ES/22ES
RF Network Analyzers
Removing the A3 Disk Drive Assembly
Tools Required
T-8 TORX screwdriver
T-10 TORX screwdriver
T-15 TORX screwdriver
small slot screwdriver
ESD (electrostatic discharge) grounding wrist strap
Removal
1. Disconnect the power cord and remove the bottom and left side covers, refer to
“Removing the Covers” on page 14-6.
2. Remove the front panel; refer to “Removing the Front Panel Assembly” on page 14-8.
3. Turn the instrument upside-down and disconnect the ribbon cable (item 1) from the
CPU board.
4. Remove the 4 screws (item 2) that secure the disk drive bracket to the side of the frame.
5. Slide the disk drive out of the instrument.
6. Remove the 4 screws (item 3) that secure the disk drive to the bracket.
Replacement