11-2 Service Guide
Error Terms 8719ET/20ET/22ET
Information on This Chapter 8719ES/20ES/22ES
RF Network Analyzers
Information on This Chapter
The analyzer generates and stores factors in internal arrays when a measurement
error-correction (measurement calibration) is performed. These factors are known by the
following names:
error terms
•E-terms
measurement calibration coefficients
The analyzer creates error terms by measuring well-defined calibration devices over the
frequency range of interest and comparing the measured data with the ideal model for the
devices. The differences represent systematic (repeatable) errors of the analyzer system.
The resulting calibration coefficients are good representations of the uncorrected
performances. For details on the various levels of error-correction, refer to the “Optimizing
Measurement Results” chapter of the analyzer user’s guide. For details on the theory of
error-correction, refer to the “Operating Concepts” chapter of the analyzer user’s guide.

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals