14-18 Service Guide
Assembly Replacement and Post-Repair Procedures 8719ET/20ET/22ET
Removing the Source Assemblies 8719ES/20ES/22ES
RF Network Analyzers
Removing the Source Assemblies
Tools Required
T-15 TORX screwdriver
5/16-inch open-end torque wrench (set to 10 in-lb)
ESD (electrostatic discharge) grounding wrist strap
Start Removal
1. Disconnect the power cord and remove the top cover; refer to “Removing the Covers” on
page 14-6.
2. Remove the front panel (refer to “Removing the Rear Panel Assembly” on page 14-14).
3. Remove the source module cover.
A58 M/A/D/S Removal
4. Disconnect the cables (item 1) and (item 2 for all but Option 400) from the M/A/D/S.
5. Remove the 4 screws (item 3) from each corner of the assembly.
Oscillator Removal
6. Remove the 3 screws (item 4) that attach the source module to the analyzer.
7. Disconnect the cables (item 1) and (item 2 for all but Option 400) from the M/A/D/S.
8. Remove the 4 screws (item 5) and (item 6) from the source module bracket. Remove the
bracket.
9. Lift the source module out of the analyzer.
10.Remove a screw (item 7) from the back of the oscillator.
11.Disconnect attaching RF cables.
A59 Source Interface Board Removal
12.Remove the 3 screws (item 4) that attach the source module to the analyzer.
13.Disconnect the cables (item 2) and (item 3 for all but Option 400) from the M/A/D/S.
14.Remove the 4 screws (item 4) and (item 5) from the source module bracket. Remove the
bracket.
15.Lift the source module out of the analyzer.
16.Remove 3 screws to detach the bottom source module cover.
17.Remove 4 screws that attach the source control board to the source module frame.
18.Place one hand on the top of the A59 board, with your thumb near the A58 M/A/D/S, to
push the board. Place your other hand on the bottom side of the A59, with your thumb
and index finger put through the drilled holes, to pull the board.

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals