Service Guide 2-35
8719ET/20ET/22ET System Verification and Performance Tests
8719ES/20ES/22ES Performance Tests
RF Network Analyzers
If the Analyzer Fails This Test
If the frequency measured is close to the specification limits (either in or out of
specification), check the time base accuracy of the counter used.
If the analyzer fails by a significant margin, the master time base probably needs
adjustment. In this case, refer to the “Frequency Accuracy Adjustment” procedure in
Chapter 3 , “Adjustments and Correction Constants.”
•See Chapter 7 , “Source Troubleshooting,” for related troubleshooting information.