6- 7
Calibrating for Increased Measurement Accuracy
Calibration Considerations
Fringe Capacitance
All open circuit terminations exhibit a phase shift over frequency due to fringe capacitance. Offset open
circuits have increased phase shift because the offset acts as a small length of transmission line. Refer to
Ta b le 6- 1 .
Table 6-1 Calibration Standard Types and Expected Phase Shift
Test Port Connector Type Standard Type Expected Phase Shift
7-mm Short 180
Type -N m al e
3.5-mm male Offset Short
3.5-mm female
2.4-mm male
2.4-mm female
Type-N female
75Type-N female
7-mm Open
Type N- m al e
3.5-mm male Offset Open
3.5-mm female
2.4-mm male
2.4-mm female
Type-N female
75 Type-N fe mal e
180
360 fl
c
-------------------------------+
0
capitance
+
0
capitance
360 fl
c
-------------------------------++