EasyManua.ls Logo

Agilent Technologies 8960 Series - Measuring Bit Error

Agilent Technologies 8960 Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
121
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started
GPRS Mobile Measurements
Measuring Bit Error
NOTE To make bit error measurements your mobile must be fitted with a Test
SIM card and have ETSI Type B connection capability.
1. Press the
CALL SETUP key.
2. Press the
More key (bottom right) to obtain the Call Parms (3 of 4) screen.
3. Press the Multislot Config (
F7) key and select 1 down, 1 up.
4. Press the Data Conn Type (
F4) key and select ETSI Type B.
5. Connect the mobile.
6. Turn the mobile on and wait for Attached in the Active Cell: field.
NOTE For mobiles that dont perform GPRS attach automatically, set the
mobile to data mode.
7. Press the Start Data Connection (
F3) key and watch for the Active Cell:
field changing to Transferring.
8. Press the
Measurement selection key.
9. Select the Bit Error measurement.
10.Press the Bit Error Setup (
F1) key.
11.Set the measurement parameters as needed for your measurement situation,
including:

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Related product manuals