cdma2000 Measurement Procedures
Measuring TDSO (Test Data Service Option) Frame Error Rate
(FER)
1. Press the CALL SETUP key.
2. If a call is currently connected, press the End Call key (
F3).
3. Press the Radio Config key (
F11).
4. Select one of the following radio configurations:
(Fwd3, Rvs3)
(Fwd4, Rvs3)
5. Press the FCH Service Option Setup key (
F12).
6. Select the Service Option for the currently selected radio configuration, then
select SO32 (+F-SCH).
7. Press the
Measurement selection key.
8. Select the TDSO Frame Error Rate measurement.
9. Press the TDSO FER Setup (
F1) key.
10.Set up the measurement parameters as needed for your measurement situation.
11.Press the
CALL SETUP key.
12.Press the Originate Call key (
F3).
13.Press the
MEASUREMENT key. Wait for the call to connect, then wait
approximately 10 seconds for the measurement to begin. A typical display is
shown below.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals