GSM Mobile Measurements
Measuring Power versus Time
1. Establish a call with the mobile.
2. Press the
Measurement selection key.
3. Select the Power vs Time measurement.
4. Press the Power vs Time Setup (
F1)key.
5. Press the Measurement Setup (
F1)key.
6. Set the measurement parameters as needed for your measurement situation,
including:
Measurement Timeout = 10 s
7. Press the Measurement Offsets (
F2) key.
Note: For statistical analysis the test set allows you to set up to 12 time markers.
These markers do not define the mask but allow you to obtain measurement
results from specified points on the mask.
8. Enter the required time offset values.
Note: Offsets are referenced to bit zero in a normal burst. To obtain results from
measurement points before bit zero, enter negative offset values.
9. Press the Close Menu (
F6) key.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals