11 Micro GC Natural Gas Analyzer
Typical Method Settings for Natural Gas Analyzers
Agilent 990 Micro GC User Manual 163
Table 35 Method for Agilent 990 Micro GC - Natural Gas Analyzer B
Method settings
*
* For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD.
PoraPLOT U CP-Sil 5 CB_6m
Carrier gas Helium Helium
Injector temperature (°C) 110 110
Injection time (ms) 40 40
Backflush time (s)
The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel.
17 -
Column temperature (°C) 60 70
Pressure (kPa) 175 150
Sample inlet temperature (°C) 110 110
Table 36 Method for Agilent 990 Micro GC - Natural Gas Analyzer B Extended
Method settings
*
* For more details of the method, see the pdf-method file available on the Natural Gas Analyzer CD.
CP-Molsieve 5A PoraPLOT U CP-Sil 5 CB_6m
Carrier gas Helium Helium Helium
Injector temperature (°C) 110 110 110
Injection time (ms) 40 40 40
Backflush time (s)
The backflush time must be tuned for each new CP-Molsieve 5A, PoraPLOT U, 4 meter CP-Sil 5 CB, and HayeSep A channel.
11 17 -
Column temperature (°C) 80 60 70
Pressure (kPa) 200 175 150
Sample inlet temperature (°C) 110 110 110

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 990 and is the answer not in the manual?

Agilent Technologies 990 Specifications

General IconGeneral
BrandAgilent Technologies
Model990
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals