8 Communications
External I/O
Agilent 990 Micro GC User Manual 123
External I/O
Connections between Micro GCs and external devices are made with the appropriate cable to the
External I/O port. The external I/O port can handle six analog inputs (input 0 to 10 Volt).
The user interface receives analog information and translates it into actions to be taken by the
local user interface, events, or data to be shown or stored in the remote user interface. In OpenLAB
EZChrom and OpenLAB ChemStation only status is visible.
Figure 53. External connections
Pin 1
EXT Analog IN 1
EXT Analog IN 2
EXT Analog IN 3
EXT Analog IN 4
EXT Analog IN 5
EXT Analog IN 6
RELAY 1 NO
RELAY 2 NO
RELAY 2 COM
START STOP NO
START STOP COM
READY NO
DGND
DGND
DGND
DGND
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
AGND
AGND
Reserved
Not used
Not used
Not used
Not used
+5 V out 500 mA
+12 V out 500 mA
+15 V out 500 mA
+12 V out 500 mA
AGND
+5 V out 500 mA
AGND
–15 V out 500 mA
Digital IN 1
Digital IN 2
EXT START IN
EXT READY IN
EXT system reset IN
Digital OUT 1
Digital OUT 2
RELAY 1 COM
READY COM