7 Replacement Procedures
Genie Filter Membrane Replacement
104 Agilent 990 Micro GC User Manual
Genie Filter Membrane Replacement
Tools required
The following tools are required to perform the replacement procedure described following
section. Allow approximately 5 to 10 minutes to complete this process.
Gloves
Torx T-20 screwdriver
Tweezers
Replacement procedure for Genie Filter membrane
1 Remove all tubing.
2 Remove the two screws with a T-20 screwdriver.
NOTE
Wear gloves to prevent contamination of the filter.