5 Micro GC Channels
PoraPlot 10 m column
68 Agilent 990 Micro GC User Manual
PoraPlot 10 m column
Table 18 PoraPlot 10 m instrument parameters
Parameter PoraPlot U 10 m Heated PoraPlot Q 10 m Heated
Column temperature 150 °C 150 °C
Injector temperature 110 °C 110 °C
Column pressure 150 kPa (21 psi) 150 kPa (21 psi)
Sample time 30 s 30 s
Injection time 40 ms 40 ms
Run time 100 s 50 s
Detector sensitivity Auto Auto
Peak 1 Composite balance Composite balance
Peak 2 Ethane 8.1 % Ethane 8.1 %
Peak 3 Propane 1.0 % Propane 1.0 %
Peak 4 i-Butane 0.14 % i-Butane 0.14 %
Peak 5 n-Butane 0.2 % n-Butane 0.2 %
mVmV PoraPlot Q 10 m heatedPoraPlot U 10 m heated
SecondsSeconds
-20
80
180
280
380
480
580
680
1
4
3
2
5
1
4
3
2
5
0102030405060
50
45
40
35
30
25
20
15
10
5
2 6 10 14 18 22 26 30 34 38 42 46 50