EasyManua.ls Logo

Agilent Technologies 990 - Method for CP-Molsieve 5 A Channel with Carrier Gas Argon

Agilent Technologies 990
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Micro GC Natural Gas Analyzer
Method for CP-Molsieve 5A Channel with Carrier Gas Argon
168 Agilent 990 Micro GC User Manual
Method for CP-Molsieve 5A Channel with Carrier Gas Argon
Typical method setting for analysis of helium and hydrogen on the CP-Molsieve 5A channel.
Table 37 Method setting for analysis of helium with Natural Gas Analyzer B Extended
Method settings CP-Molsieve 5A
Carrier gas Argon
Injector temperature (°C) 110
Injection time (ms) 40
Backflush time (s)
*
* The backflush time must be tuned for each new CP-Molsieve 5A channel.
11
Column temperature (°C) 80
Pressure (kPa) 200
Sample inlet temperature (°C) 110
Invert signal Yes

Table of Contents

Related product manuals