2 Instrument Overview
Ambient Pressure
Agilent 990 Micro GC User Manual 29
Ambient Pressure
The Micro GC automatically shuts down if the ambient pressure is greater than 120 kPa for a
standard Micro GC, or is greater than 110 kPa for the Energy meter version.
Ambient Temperature
The operating temperature for the Micro GC is 0-50 °C. The Micro GC automatically shuts down if
the ambient temperature exceeds 70 °C.
Maximum Operation Altitude
Certified up to 2,000 meters above sea level.
Storage requirements
•Humidity: 10 % to 95 % RH (noncondensing)
•Temperature: –40 to +70 °C