5 Micro GC Channels
Volamine Column
Agilent 990 Micro GC User Manual 71
Volamine Column
Table 21 CP-Volamine instrument parameters
Parameter CP-Volamine 15m Heated
Column temperature 50 °C
Injector temperature 50 °C
Column pressure 150 kPa (21 psi)
Sample time 30 s
Injection time 40 ms
Run time 60 s
Detector sensitivity Auto
Peak 1 Nitrogen Balance
Peak 2 Ammonia 100 ppm
0
20
40
60
80
100
0.5 0.55 0.6 0.65 0.7
1
2
Minute
µV