7 Replacement Procedures
Replacement procedure for Genie Filter membrane
Agilent 990 Micro GC User Manual 105
3 Separate the two parts. Considerable force may be required to separate the two parts of the
filter.
4 Use tweezers to remove the attached white membrane.
5 Place a new membrane at the center of the part, and replace the top.
6 Replace the two screws, and tighten with a T-20 screwdriver.