Agilent InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
s1
7
Serial Decode
Serial Decode Options 123
Lister 124
Searching Lister Data 126
Triggering on
Serial Data
In some cases, such as when triggering on a slow serial signal (for
example, I2C, SPI, CAN, LIN, etc.) it may be necessary to switch from the
Auto trigger mode to the Normal trigger mode to prevent the oscilloscope
from Auto- triggering and stabilize the display. You can select the trigger
mode by pressing the [Mode/Coupling] key, then the Mode softkey.
Also, the threshold voltage level must be set appropriately for each source
channel. The threshold level for each serial signal can be set in the Signals
Menu. Press the [Serial] key, then the Signals softkey.
Serial Decode Options
Agilent's hardware- accelerated serial decode options can be installed when
the oscilloscope is manufactured, or added later. The following serial
decode licenses are available:
With the DSOX3AUTO license, you can decode CAN (Controller Area
Network) and LIN (Local Interconnect Network) serial buses. See:
"CAN Serial Decode" on page 339.
"LIN Serial Decode" on page 347.
With the DSOX3FLEX license, you can decode FlexRay serial buses. See
"FlexRay Serial Decode" on page 355.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 MSO-X 3012A and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 MSO-X 3012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 MSO-X 3012A
CategoryTest Equipment
LanguageEnglish

Related product manuals