Trigger Mode/Coupling 11
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 183
The external trigger input impedance is 1M Ohm. This lets you use passive
probes for general- purpose measurements. The higher impedance
minimizes the loading effect of the oscilloscope on the device under test.
To set the EXT TRIG IN units and probe attenuation:
1 Press the [Mode/Coupling] key in the Trigger section of the front panel.
2 In the Trigger Mode and Coupling Menu, press the External softkey.
3 In the External Trigger Menu, press the Units softkey to select between:
Volts — for a voltage probe.
Amps — for a current probe.
Measurement results, channel sensitivity, and the trigger level will
reflect the measurement units you have selected.
4 Press the Probe softkey; then, turn the entry knob to specify the probe
attenuation.
The attenuation factor can be set from 0.1:1 to 1000:1 in a 1-2- 5
sequence.
The probe attenuation factor must be set properly for measurements to
be made correctly.
CAUTION
Maximum voltage at oscilloscope external trigger input
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
1 M ohm input: For steady-state sinusoidal waveforms derate at 20 dB/decade above
57 kHz to a minimum of 5 Vpk
With N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
With 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals