Triggers 10
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 161
 
9 Press the <Hold softkey, and turn the Entry knob to select the hold 
time.
Video Trigger
Video triggering can be used to capture the complicated waveforms of 
most standard analog video signals. The trigger circuitry detects the 
vertical and horizontal interval of the waveform and produces triggers 
based on the video trigger settings you have selected.
The oscilloscope's MegaZoom IV technology gives you bright, easily viewed 
displays of any part of the video waveform. Analysis of video waveforms is 
simplified by the oscilloscope's ability to trigger on any selected line of the 
video signal.
1 Press the [Trigger] key.
It is important, when using a 10:1 passive probe, that the probe is correctly compensated. 
The oscilloscope is sensitive to this and will not trigger if the probe is not properly 
compensated, especially for progressive formats.