Measurements 14
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 219
"Maximum" on page 219
"Minimum" on page 219
"Amplitude" on page 219
"Top" on page 219
"Base" on page 220
"Overshoot" on page 221
"Preshoot" on page 222
"Average" on page 222
"DC RMS" on page 223
"AC RMS" on page 223
"Ratio" on page 225
Peak-Peak
The peak- to- peak value is the difference between Maximum and Minimum
values. The Y cursors show the values being measured.
Maximum
Maximum is the highest value in the waveform display. The Y cursor
shows the value being measured.
Minimum
Minimum is the lowest value in the waveform display. The Y cursor shows
the value being measured.
Amplitude
The Amplitude of a waveform is the difference between its Top and Base
values. The Y cursors show the values being measured.
Top
The Top of a waveform is the mode (most common value) of the upper
part of the waveform, or if the mode is not well defined, the top is the
same as Maximum. The Y cursor shows the value being measured.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals