Agilent InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
s1
22
CAN/LIN Triggering and Serial Decode
Setup for CAN Signals 335
CAN Triggering 337
CAN Serial Decode 339
Setup for LIN Signals 343
LIN Triggering 345
LIN Serial Decode 347
CAN/LIN triggering and serial decode requires Option AUTO or the
DSOX3AUTO upgrade.
Setup for CAN Signals
Setup consists of connecting the oscilloscope to a CAN signal, using the
Signals Menu to specify the signal source, threshold voltage level, baud
rate, and sample point.
To set up the oscilloscope to capture CAN signals, use the Signals softkey
which appears in the Serial Decode Menu:
1 Press [Label] to turn on labels.
2 Press [Serial].
3 Press the Serial softkey, turn the Entry knob to select the desired slot
(Serial 1 or Serial 2), and press the softkey again to enable decode.
4 Press the Mode softkey; then, select CAN trigger type.
5 Press the Signals softkey to open the CAN Signals Menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals