4 Connect a BNC cable (40 inches maximum) from the TRIG OUT 
connector on the rear panel to the BNC barrel connector.
5 Press the [Utility] key; then, press the Service softkey.
6 Begin the Self Cal by pressing the Start User Cal softkey.
To perform hardware self test
Pressing  [Utility] > Service > Hardware Self Test performs a series of internal 
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
• After experiencing abnormal operation.
• For additional information to better describe an oscilloscope failure.
• To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the 
oscilloscope's functionality. Hardware Self Test is designed to provide an 
80% confidence level that the oscilloscope is operating properly.
To perform front panel self test
Pressing  [Utility] > Service > Front Panel Self Test lets you test the front panel 
keys and knobs as well as the oscilloscope display.
Follow the on- screen instructions.
To display oscilloscope information
Press  [Help] > About Oscilloscope to display information about your 
oscilloscope:
• Model number.
• Serial number.
• Bandwidth.
• Module installed.
• Software version.
• Installed licenses. See also "Loading Licenses and Displaying License 
Information" on page 323.