Y Cursors Y cursors are horizontal dashed lines that adjust vertically and can be 
used to measure Volts or Amps, dependent on the channel Probe Units 
setting, or they can measure ratios (%). When math functions are used as 
a source, the measurement units correspond to that math function.
The Y1 cursor is the short- dashed horizontal line and the Y2 cursor is the 
long- dashed horizontal line.
The Y cursors adjust vertically and typically indicate values relative to the 
waveform's ground point, except math FFT where the values are relative 
to 0 dB.
In XY horizontal mode, the Y cursors display channel 2 values (Volts or 
Amps).
When active, the Y1 and Y2 cursor values for the selected waveform 
source are displayed in the softkey menu area.
The difference between Y1 and Y2 (ΔY) is displayed in the Cursors box in 
the right- side information area.
To make cursor measurements
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Press the [Cursors] key.
The Cursors box in the right- side information area appears, indicating 
that cursors are "on". (Press the [Cursors] key again when you want to 
turn cursors off.)
3 In the Cursors Menu, press Mode; then, select the desired mode:
• Manual — ΔX, 1/ΔX, and ΔY values are displayed. ΔX is the difference 
between the X1 and X2 cursors and ΔY is the difference between the 
Y1 and Y2 cursors.