CAN/LIN Triggering and Serial Decode 22
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 345
 
12 Press the Standard softkey; then, turn the Entry knob to select the LIN 
standard you are measuring (LIN 1.3 or LIN 2.0).
For LIN 1.2 signals, use the LIN 1.3 setting. The LIN 1.3 setting assumes 
the signal follows the "Table of Valid ID Values" as shown in section A.2 
of the LIN Specification dated December 12, 2002. If your signal does 
not comply with the table, use the LIN 2.0 setting.
13 Press the Sync Break softkey and select the minimum number of clocks 
that define a sync break in your LIN signal.
LIN Triggering
To set up the oscilloscope to capture a LIN signal, see "Setup for LIN 
Signals" on page 343.
LIN triggering can trigger on the rising edge at the Sync Break exit of the 
LIN single- wire bus signal (that marks the beginning of the message 
frame), the Frame ID, or the Frame ID and Data.
A LIN signal message frame is shown below:
1 Press  [Trigger].
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob 
to select the serial slot (Serial 1 or Serial 2) on which the CAN signal 
is being decoded.
60%
70%
80%
Sample Point
One Bit
Sync
Break
Identifier
Break
Sync Break Exit
Sync
Field
Checksum
Field
Data
Fields