Vertical Controls 3
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 65
 
When AC coupling is selected (see "To specify channel coupling" on 
page 63) or excessive voltage is applied to the input, the oscilloscope 
automatically switches to 1M Ohm mode to prevent possible damage.
• 1M Ohm — is for use with many passive probes and for 
general- purpose measurements. The higher impedance minimizes the 
loading effect of the oscilloscope on the device under test.
This impedance matching gives you the most accurate measurements 
because reflections are minimized along the signal path.
See Also • For more information on probing, visit: 
"www.agilent.com/find/scope_probes"
• Information about selecting a probe can be found in document number 
"Agilent Oscilloscope Probes and Accessories Selection Guide (part 
number 5989- 6162EN)", available at "www.agilent.com".
To specify bandwidth limiting
1 Press the desired channel key.
2 In the Channel Menu, press the BW Limit softkey to enable or disable 
bandwidth limiting.
When bandwidth limit is on, the maximum bandwidth for the channel is 
approximately 20 MHz. For waveforms with frequencies below this, turning 
bandwidth limit on removes unwanted high frequency noise from the 
waveform. The bandwidth limit also limits the trigger signal path of any 
channel that has BW Limit turned on.
To change the vertical scale knob's coarse/fine adjustment setting
1 Push the channel's vertical scale knob (or press the channel key and 
then the Fine softkey in the Channel Menu) to toggle between fine and 
coarse adjustment of the vertical scale.
When  Fine adjustment is selected, you can change the channel's vertical 
sensitivity in smaller increments. The channel sensitivity remains fully 
calibrated when Fine is on.