MIL-STD-1553/ARINC 429 Triggering and Serial Decode 26
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 397
Aliased data is highlighted in pink. When this happens, decrease the
horizontal time/div setting and run again.
Searching for MIL-STD-1553 Data in the Lister
The oscilloscope's search capability lets you search for (and mark) certain
types of MIL- STD- 1553 data in the Lister. You can use the [Navigate] key
and controls to navigate through the marked rows.
1 With MIL- STD- 1553 selected as the serial decode mode, press [Search].
2 In the Search Menu, press the Search softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the
MIL- STD-1553 signal is being decoded.
3 Press Search; then, select from these options:
Data Word Start — finds the start of a Data word (at the end of a valid
Data Sync pulse).
Command/Status Word Start — finds the start of Command/Status word
(at the end of a valid C/S Sync pulse).
Remote Terminal Address — finds the Command/Status word whose RTA
matches the specified value. The value is specified in hex.
When this option is selected, the RTA softkey becomes available and
lets you select the hex Remote Terminal Address value to find.
Remote Terminal Address + 11 Bits finds the RTA and the remaining 11
bits that match the specified criteria.
When this option is selected, these softkeys become available:
The RTA softkey lets you select the hex Remote Terminal Address
value.
The Bit Time softkey lets you select the bit time position.
The 0 1 X softkey lets you set the bit time position value as a 1, 0,
or X (don't care).
Parity Error — finds (odd) parity bits that are incorrect for the data in
the word.
Sync Error — finds invalid Sync pulses.
Manchester Error — finds Manchester encoding errors.
For more information on searching data, see "Searching Lister Data" on
page 126.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals